Oxygen concentrators were invented in the early 1970s for use in the human home health-care industry for patients with various medical conditions, such as chronic obstructive pulmonary disease (COPD), emphysema, bronchitis, etc. Manufacturing output of these devices increased later that decade. Prior to that, home medical oxygen therapy required the use of heavy high-pressure oxygen cylinders or small cryogenic liquid oxygen systems. In the mid-1980s, Medicare switched to a flat monthly rate for home oxygen, causing the industry to rapidly embrace concentrators as a way to “control costs.” Oxygen concentrators became the preferred and most common means of delivering home oxygen. This brief history of oxygen concentrators is important information for the veterinary medical professional, as it closely follows suit with our industry.
Choosing the right oxygen concentrator system for your practice
Oxygen concentrators can be a safe and attractive alternative to using high-pressure oxygen cylinders and/or liquid oxygen. Overall, users can lower their clinic’s operating costs as a result of not using equipment subject to price increases, missed deliveries, or being locked into contracts with their oxygen supplier. On the other hand, it is important to note that if an oxygen concentrator system becomes compromised (e.g. loss of electrical power), the facility must have a backup oxygen supply available at all times, whether it is a high-pressure oxygen cylinder or liquid oxygen.
To successfully size a precise oxygen concentrator system for your overall requirements, it is recommended you deal with an established and reputable manufacturer. With that said, the following questions regarding existing equipment should be reviewed with the respective manufacturer to address oxygen requirements at your facility:
1) How many anesthesia machines will you be operating at any given time? Each unit requires between 1 to 2 liters per minute (LPM) (oxygen-flow rate).
2)Will you be operating a ventilator in conjunction with an anesthesia machine(s)? Each ventilator requires on average of 4 LPM at 50 psi (output pressure), so it’s important to know how many ventilators you’ll be using.
3) How many intensive care units and/or oxygen recovery cages will you be operating at any given time? Each unit requires on average upward of 15 LPM to initially “flush” the system and then 5 LPM “to maintain” it. It is critically important to verify with the respective manufacturer/vendor the LPM oxygen-flow rate setting required for this type of equipment.
4) Regarding the first three questions, what is the total number of items/equipment you would be operating simultaneously? By determining this, you will then know the exact liter-per-minute requirement at your facility at a given time.
5) Do you operate any other equipment requiring oxygen (incubators, nebulization, etc.)? Although this would be a relatively small liter-per-minute figure, it is important to know what additional oxygen-flow rate you will need based on the items at your facility requiring oxygen.
6) In addition to excellent ventilation, every oxygen concentrator system requires adequate room/space in which to operate properly. Is this available at your facility? This consideration is sometimes overlooked, and can have a significant effect on the overall performance and longevity of your system. Please note room size is not the underlying issue, but rather the airflow through the room. You need to have enough incoming air to ensure the system is not choked off and have enough outflowing air to ensure the room does not get warmer due to heat from the compressor. Further, care must be given to ensure the space does not become a nitrogen-enriched environment (waste gas).
7) Will you be using a centralized oxygen pipeline/manifold system to deliver oxygen from the oxygen concentrator to all respective equipment? If so, keep in mind you will require a minimum of 20 PSI (output pressure). On the other hand, if you are only using one or two anesthesia machines with a direct connection to these units, a small 10-LPM oxygen concentrator would be adequate. That said, it is recommended the unit also have at least 20 PSI (output pressure). The output pressure is important to operate the flush valve on the anesthesia machine(s).
8) At what elevation (feet or meters above sea level) is your clinic or hospital? An elevation of more than 457 m (1500 ft) above sea level may affect the operation of your system from a flow rate and output pressure standpoint.
Veterinary clinics, like many other medical care facilities, involve the supply of oxygen generation. The purpose of oxygen generators for veterinarians is for oxygen supply to be available quickly and efficiently, disposing of any risk to the animals. 95% oxygen generators are used to stabilize animals both during and after surgical procedures. Installations for our on-site oxygen generation systems range from small clinics to large hospitals. Typical oxygen generators are installed indoors, with feed lines running into the operating rooms – so it does not take up the minimal space in most surgery rooms.